Refractometer designed for InLine applications and to adapt into a process connection flange. The Instrument
Extremely sensitive and precise measurement of process liquids and stable measuring values.
The refractive index can be calibrated in any required concentration unit and is available as a digital signal at RS485 standards and as a standard 4...20mA output signal.
Instrument can be fitted directly into the tank wall, so no sampling system required!
The ProcanND-TW01 is connected to the ProcanMaster control unit RCU-01. The microprocess computer of this control unit takes care for the calculation of the concentration and generates the digital or analogue outputs.

Specifications:
| Measuring range | max. nD 1,20000...1,65000 |
| Precision | <0,1% of range |
| Sensitivity | <0,01% of range |
| Drift | none |
| Speed of response | T90 < 2s |
| Temperature compensation | is included in the calibration and implemented over range of interest and +\- 5 oC of default process temperature |
| Process temperature | from 5 ... 65 oC |
| Power supply | via ProcanMaster control unit RCU-01 |
| Output signals |
measuring value: 4..20 mA, temp.value: pt100 standard |
Access Units: different types of process connections are available. Flanges are to weld into the tank wall. Enclosure: Stainless Steel 316, IP64
Materials in contact process liquid: SS316, Teflon, Viton, Saphire
Contact us for more information
Steuma Instruments
Kortestraat 10a
4818LT Breda
The Netherlands
Tel. +31 76 5220901
Fax. +31 76 5205865
Email info@steuma.com